Medical Standard IEC60601-1 certified, ATX Power Supply |
|
Double and rainforced insulation type, so that Medical standard approved commercial insulating transformer is unnecessary. - Fuses are mounted on both L/N line -Low leakage current: under 0.3mA at AC 264V input. |
|
Continuous
301W |
|
Peak
500.5W |
|
|
*Output Harnesses are optional. |
*Please use Metric Screw Threads, ("M3" screws) for PSU mounting.(Inch Screw Threads cannot be used.) |
Function :For description of function, click each icon
Output specification
Output connector : For description of function, click each icon
Standard, specifications: For standard and specifications, click each item.
Compatible case size |
PS2 |
Outline drawing |
150Wx86Hx139D |
Weight |
1.80 kg typ |
Input Voltage |
AC |
85-264V |
Efficiency |
AC |
73% typ(AC100V)/77% typ(AC240V) |
Power factor |
99% typ(AC100V)/97% typ(AC240V) |
Input changeover system |
Worldwide |
Safety |
UL60601-1, CSA60601-1(c-UL), IEC60601-1, UL60950-1, CSA60950-1(c-UL), CCC, CE marking(IEC62368-1) |
High frequency current regulation |
IEC61000-3-2 (Ed. 2.1) class D EN61000-3-2 (A14)class D |
EMI standard (Input feedback noise) |
Input feedback noise VCCI class B FCC standard class B EN55022 class B |
CE marking |
Low voltage and EMI directive and RoHS2.0 |
EMS immunity |
Electrostatic discharge to meet EN61000-4-2. Radioactive radio-frequency electromagnetic fields to meet EN61000-4-3 Fast transient burst to meet EN61000-4-4 Lightning surge immunity to meeet EN61000-4-5 Immunity to conducted disturbances, Conductive radio-frequency electromagnetic field to meet EN61000-4-6 Power frequency magnetic field immunity to meet EN61000-4-8 Voltage dips and fluctuation to meet EN61000-4-11 |
MTBF |
|
Operating temperature |
0 to 60 degrees C (Derating operation needed at 45 to 60 degrees C) |
Leakage current |
0.12mA max(AC100V)/0.3mA max(AC264V) |
Cooling systems |
Forced air cooling |
Insulation |
|
Output ground connection |
|
Warranty |
Three years after delivery. However, if defects belong us, the defective unit shall be repaired or replaced at our cost. |